LTOBR4600 REV. 004 02/13/2014
Spatial Resolution (two-point)
1
:
Equals 2-pt spatial resolution
Integrated Return Loss Characteristics:
Integrated Insertion Loss Characteristics:
Distributed Sensing
7,12,13
:
Time vs. wavelength range
Long Range (2 km) Scan Time
Specifications are for single-mode operation.
For multimode operation, specifications are nominal.
1 Over entire length range.
2 Ranges are nominal.
3 Accuracy maintained by an internal NIST-traceable HCN
gas cell.
4 For the 2000 m option, return loss dynamic range is 60 dB.
5 With integration width of 0.5 m
6 The insertion loss dynamic range is the one-way loss
that can be suffered before the scatter level of standard
SMF (~ -100 dB/mm) is lower than the noise floor
(~ -118 dB/mm).
7 Distributed sensing uses Rayleigh spectral shift method
and is relative to reference scan.
8 Spatial resolutions listed are ideal to get the temperature
and strain resolutions listed; they are not minimums or
maximums.
9 Temperature and strain resolutions are calculated from
spectral shift of Rayleigh scatter using 1 GHz = 0.8 C=
6.58 µStrain. [Othonos and K. Kalli, Fiber Bragg
Gratings (Artech House, Boston, 1999)].
10 Combined scan and analysis time in high-resolution
mode. Times are for 30 m scan mode.
11 Times are with laser tuning speed set at 100 nm/s.
12 Extended range mode and distributed sensing are
upgrade options
13 Maximum standard sensing length is 70m. Limited
sensing is available in extended range mode. Contact
your Luna representative for more information.